| Model Number |
INFORM II |
| Medical Device Problem Code |
Circuit Failure (1089)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/08/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The initial reporter complained of an issue with an accu-chek inform ii base unit.The customer alleged the base unit was not charging and the charging contacts of the base unit appeared "scorched" and "burnt.".
|
| |
|
Additional Manufacturer Narrative
|
|
The base unit was requested for investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
The customer did not return the product for investigation.Since no product was returned, the cause of the event could not be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received for investigation.A visual examination was performed and corrosion was observed on one of the charging contacts.This corrosion can affect the charging capabilities of the base unit.The investigation determined the corrosion was consistent with a customer handling issue (improper cleaning).The investigation did not identify a product problem.
|
| |
|
Search Alerts/Recalls
|