Model Number BEA25-120/I16-40
Medical Device Problem Code Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 09/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft and an afx limb stent graft to treat an abdominal aortic aneurysm (aaa).Approximately (1) one-year post initial procedure, an endoleak (indeterminate origin) was identified.The bifurcated stent graft also showed as collapsed.The patient's current condition is unknown, and it is unknown if the physician plans to re-intervene.
 
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.Device remains implanted.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good-faith efforts to retrieve a reported adverse event/incident-related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident could not be completed.No medical records nor medical imaging relevant to the reported adverse event/incident were received by endologix.The user facility was unable to provide this information.Procedure-related harms, device, user, procedure, or anatomy-relatedness of this complaint could not be determined with the medical records available for review.The final patient status was reported as being fine with no complaints.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.
 
Event or Problem Description
The patient was initially implanted with an afx2 bifurcated stent graft and an afx limb stent graft to treat an abdominal aortic aneurysm (aaa).Approximately (1) one-year post initial procedure, an endoleak (indeterminate origin) was identified.The bifurcated stent graft also showed as collapsed.The patient's current condition is unknown, and it is unknown if the physician plans to re-intervene.Additional information: reintervention was completed with the implant of an afx vela infrarenal.The final patient status was reported as being fine with no complaints.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15620290
Report Number2031527-2022-00248
Device Sequence Number6546654
Product Code MIH
UDI-Device Identifier00818009014719
UDI-Public(01)00818009014719(17)230616
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 09/28/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/16/2023
Device Model NumberBEA25-120/I16-40
Device Lot Number2362979-005
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2022
Supplement Date Received by Manufacturer11/06/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date12/21/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/07/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX LIMB STENT GRAFT 2044842-004.
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient SexMale
-
-