• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number N5C8305C
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received and is currently awaiting evaluation.Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported an unknown quantity (also reported as potentially twelve) of low recirculation volume apd set with cassettes were damaged.There was a "slice" observed in the tubing during set up.There was no damaged noted to the overpouch of the cassette and no other damage was visible upon inspection of the product.
 
Additional Manufacturer Narrative
B5: there was no patient involvement.No additional information is available.H10: twelve devices were received for evaluation.Three of the devices were received in an open pouch and nine were received in unopened pouches.During visual inspection, scuff/scratches were observed on the drain line tubing and supply line tubing next to the tack weld.The sets underwent functional testing, including leak, clear passage and clamp functioning testing, and the sets performed according to product specifications.No leaks observed in any of the devices.The scuff/scratches are cosmetic in nature and do not affect the functionality of the device.Therefore, there was no risk to the patient as the scuff/scratches on the tubing did not result in a breach in the sterile pathway.The reported condition was verified.The cause of the condition was determined to be manufacturing related, caused by tubing grippers during the automated assembly process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE LOW RECIRCULATION VOLUME APD SET WITH CASSETTE
Common Device Name
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15620546
Report Number1416980-2022-05533
Device Sequence Number6385835
Product Code FKX
UDI-Device Identifier00085412087672
UDI-Public(01)00085412087672
Combination Product (Y/N)N
Initial Reporter StateDC
Initial Reporter CountryUS
PMA/510(K) Number
K102936
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 11/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberN5C8305C
Device Lot NumberH21F07031
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/05/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/19/2022
Supplement Date Received by Manufacturer10/24/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date11/16/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-