• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LIMITED FLEXITRUNK INFANT NASAL TUBING; BZO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LIMITED FLEXITRUNK INFANT NASAL TUBING; BZO Back to Search Results
Model Number BC192-05
Medical Device Problem Code Material Split, Cut or Torn (4008)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) are currently in the process of retrieving further information regarding this complaint.We will provide a follow up report upon the completion of our investigation.
 
Event or Problem Description
A distributor in india reported via a fisher and paykel (f&p) healthcare representative, that bc192 flexitrunk infant nasal tubing was found broken before patient use.There was no patient involvement.
 
Event or Problem Description
A distributor in india reported via a fisher and paykel (f&p) healthcare representative, that bc192 flexitrunk infant nasal tubing was found broken before patient use.There was no patient involvement.
 
Additional Manufacturer Narrative
(b)(4).Method: the complaint bc192 flexitrunk infant nasal tubing was not returned to the fisher & paykel healthcare (f&p) for evaluation.Our investigation is therefore based on the information and photograph provided by the customer.Results: the visual inspection of the provided photograph revealed the tubing is damaged.The film is wrinkled and torn.Conclusion: without the complaint device, we are unable to determine the cause of the reported failure.However, it is likely to have been caused by pulling of the tubing.All flexitrunk nasal tubings are visually inspected, and pressure tested before leaving the production line and those that fail are rejected.This suggests that the damage on the subject nasal tubing occurred after it was released for distribution.Our user instructions the accompany the flexitrunk infant nasal tubing state: - "use caution when positioning the infant interface.Avoid excessive pull forces, sharp objects and tubing holders.Damage to the tubing may cause loss of pressure and require immediate replacement." a caution insert is included in the packaging to remind the user to take extra care when handling the flexitrunk.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXITRUNK INFANT NASAL TUBING
Common Device Name
BZO
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LIMITED
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
faranak gomarooni
17400 laguna canyon road
suite 300
irvine, CA 92618
9494534002
MDR Report Key15620679
Report Number9611451-2022-00975
Device Sequence Number6184358
Product Code BZO
UDI-Device Identifier09420012413406
UDI-Public(01)09420012413406(10)2102117410(11)220401
Combination Product (Y/N)N
Initial Reporter CountryIN
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative,Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/26/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberBC192-05
Device Catalogue NumberBC192-05
Device Lot Number2102117410
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer12/02/2022
Initial Report FDA Received Date10/17/2022
Supplement Report FDA Received Date12/02/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-