STRYKER ENDOSCOPY-SAN JOSE INJECTOR II NEEDLE CARTRIDGE; NEEDLE, SUTURING, DISPOSABLE Back to Search Results
Model Number CAT01857
Medical Device Problem Code Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/12/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the needle broke during the procedure.The needle was retrieved.
 
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: tip of the injector needle broke probable root cause: design - design stackup interference, too much clearance between catridge and loaded suture - catridge tip goemetry not sufficient for grabbing suture - cartridge design does not withstand attempted passing forces, material too weak - handle tip design does not withstand attempted passing forces, too weak - handle tip design insufficient for suture retention after passing process - handle tip not manufactured to specification - cartridge tip not manufactured to specification application - incorrect cartridge assembly - reuse of catridge - damage to cartridge during handling - difficult surrounding tissue or anatomy - improper loading of suture the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.81.
 
Event or Problem Description
It was reported that the needle broke during the procedure.The needle was retrieved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INJECTOR II NEEDLE CARTRIDGE
Common Device Name
NEEDLE, SUTURING, DISPOSABLE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer (Section G)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
Manufacturer Contact
lucas wolski
5900 optical court
san jose, CA 95138
4087542000
MDR Report Key15621024
Report Number0002936485-2022-00596
Device Sequence Number5978954
Product Code GAB
UDI-Device Identifier07613252633105
UDI-Public07613252633105
Combination Product (Y/N)N
Initial Reporter CountryCI
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/30/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCAT01857
Device Catalogue NumberCAT01857
Device Lot Number006901
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer09/26/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date11/30/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1