MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ACUITY X4 STRAIGHT; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ACUITY X4 STRAIGHT; IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT) Back to Search Results
Model Number 4671
Device Problem Pocket Stimulation (1463)
Patient Problems Device Overstimulation of Tissue (1991); Discomfort (2330)
Event Date 10/06/2022
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient with this left ventricular (lv) lead experienced severe diaphragmatic stimulation during lv pacing.Moreover, all vectors were tested to minimize diaphragmatic stimulation, however all pacing vectors were still causing patient discomfort.The lead was surgically abandoned, and new lead was implanted.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACUITY X4 STRAIGHT
Type of Device
IMPLANTABLE PULSE GENERATOR, PACEMAKER (NON-CRT)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key15621233
MDR Text Key301904896
Report Number2124215-2022-41967
Device Sequence Number1
Product Code LWP
UDI-Device Identifier00802526524516
UDI-Public00802526524516
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
P010012/S398
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/21/2023
Device Model Number4671
Device Catalogue Number4671
Device Lot Number852029
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexFemale
-
-