Model Number BEA28-70/I16-30
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Codes Aneurysm (1708); Failure of Implant (1924)
Date of Event 10/06/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The devices involved in this event will not be returned for evaluation and remain implanted in the patient.Patient medical records and imaging studies will be requested for further evaluation by a clinical specialist.If additional information pertinent to the incident is obtained, a follow-up report will be submitted.Device iteration is afx2.
 
Event or Problem Description
During the initial implant of an afx2 bifurcated stent graft and afx infrarenal aortic extension, intraoperative type ia, iiia and iiib endoleaks were reported.The patient first had a gore (non-endologix) excluder device implanted in the common iliac artery.The afx2 bifurcated stent graft and afx infrarenal aortic extension were then implanted.An angiogram after balloon touch-up showed a type ia endoleak, and therefore a second gore(non-endologix) excluder was implanted to resolve this leak.A type iiia endoleak was then confirmed from the left side, and thus a third gore (non-endologix) excluder was added to resolve this leak.Another endoleak was observed above the terminal aorta, and it was suspected as a type iiib endoleak.Additional intervention was not performed at the time, and the physician elected to monitor the patient.Note: this procedure is outside the indications of use (off-label) due to adjunctive devices not compatible with the afx system per device ifu.This event was previously reported per mfr.Report # 2031527-2020-00080.Approximately two and a half (2.5) years post initial procedure, the patient presented at routine follow-up with a possible type ii endoleak (of the lumbar arteries) or a type iiia endoleak per ct (computed tomography) scan.The physician elected to treat the patient by implanting one (1) afx vela infrarenal and one (1) gore (non-endologix) excluder cuff on (b)(6) 2022.During this secondary intervention, it was confirmed that a type iiia endoleak was present.While the endoleak was not completely resolved, the operation was concluded as its volume had decreased.The patient will continue to be monitored.
 
Event or Problem Description
During the initial implant of an afx2 bifurcated stent graft and afx infrarenal aortic extension, intraoperative type ia, iiia and iiib endoleaks were reported.The patient first had a gore (non-endologix) excluder device implanted in the common iliac artery.The afx2 bifurcated stent graft and afx infrarenal aortic extension were then implanted.An angiogram after balloon touch-up showed a type ia endoleak, and therefore a second gore(non-endologix) excluder was implanted to resolve this leak.A type iiia endoleak was then confirmed from the left side, and thus a third gore (non-endologix) excluder was added to resolve this leak.Another endoleak was observed above the terminal aorta, and it was suspected as a type iiib endoleak.Additional intervention was not performed at the time, and the physician elected to monitor the patient.Note: this procedure is outside the indications of use (off-label) due to adjunctive devices not compatible with the afx system per device ifu.This event was previously reported per mfr.Report # 2031527-2020-00080.Approximately two and a half (2.5) years post initial procedure, the patient presented at routine follow-up with a possible type ii endoleak (of the lumbar arteries) or a type iiia endoleak per ct (computed tomography) scan.The physician elected to treat the patient by implanting one (1) afx vela infrarenal and one (1) gore (non-endologix) excluder cuff on 11 october 2022.During this secondary intervention, it was confirmed that a type iiia endoleak was present.While the endoleak was not completely resolved, the operation was concluded as its volume had decreased.The patient will continue to be monitored.Additional information: the physician elected to further treat the patient by implanting an aorfix (non-endologix) stent graft on 31 october 2022.While the volume of the endoleak decreased, it was not completely resolved.Patient will continue to be monitored.Clinical assessment confirmed that a type ii endoleak (non-device-related) and significant aneurysm sac enlargement were also present.
 
Additional Manufacturer Narrative
The reported adverse event/incident was investigated in alignment with endologix operating procedures and work instructions.Where possible, it is endologix practice to make at least three good faith efforts to retrieve a reported adverse event/incident related device as well as medical records and medical imaging.An evaluation of the manufacturing record was completed.A review of the part number/lot number combination needed for device identification shows the device to have been properly manufactured and released in accordance with the device master record.The review confirms there were no manufacturing or processing non-conformities identified that would contribute to the reported adverse event/incident.An evaluation of the device could not be completed.The device was not returned to endologix for evaluation because it remains implanted.A clinical evaluation of the adverse event/incident was completed.An examination of medical records and/or medical imaging received by endologix shows the reported type iiia endoleak (of the aortic component) is unconfirmed; while there was a clear mid-aortic endoleak, it could not be confirmed as a type iiia endoleak.The reported additional endovascular procedure is confirmed.This is moderately consistent with the reported adverse event/incident.The clinical evaluation also shows reasonable evidence to suggest a non-device-related type ii endoleak (of the lumbar arteries) and aneurysm enlargement (of 8.1mm) occurred that were not included in the event as reported.These findings were discovered during an examination of the 30-month post implant ct scan.There was a non-endologix proximal extension placed at initial procedure prior to implant of any afx products (off-label condition); this could have contributed to the reported event, but that could not be conclusively determined.In addition, the pre-index proximal aortic neck measured 32.4mm (device ifu requirement is 18-32mm).Procedure-related harms, device, user, procedure, or anatomy relatedness of this event could not be determined with the medical records available for review.The final patient status was not reported.No additional investigation of this reported adverse event/incident is planned.However, should additional information relevant to the investigation outcome become available, a follow-up report will be submitted.Endologix will continue to monitor this and similar adverse events/incidents.Device iteration is afx2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFX2
Common Device Name
BIFURCATED STENT GRAFT
Manufacturer (Section D)
ENDOLOGIX
2 musick
irvine CA 92618
Manufacturer (Section G)
ENDOLOGIX
2 musick
,
irvine CA 92618
Manufacturer Contact
gary kirchgater
2 musick
,
irvine, CA 92618
8009832284
MDR Report Key15621561
Report Number2031527-2022-00250
Device Sequence Number5978979
Product Code MIH
UDI-Device Identifier00818009014603
UDI-Public(01)00818009014603(17)221024
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P040002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall,Other
Type of Report Initial,Followup
Report Date (Section B) 10/11/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/24/2022
Device Model NumberBEA28-70/I16-30
Device Lot Number2301607-002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/11/2022
Supplement Date Received by Manufacturer12/12/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured11/20/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0007-2019
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
AFX INFRARENAL AORTIC EXTENSION, SN (B)(4); GORE (NON -ENDOLOGIX) EXCLUDER CUFF PLA280300J; GORE (NON -ENDOLOGIX) EXCLUDER CUFF PLA280300J; GORE (NON -ENDOLOGIX) EXCLUDER CUFF PLA280300J
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
-
-