• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION PT; WIRE, GUIDE, CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION PT; WIRE, GUIDE, CATHETER Back to Search Results
Model Number 2783
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Patient age at time of event: 18 years or older.
 
Event or Problem Description
It was reported that guidewire break occurred.A 300cm pt2 moderate support guidewire was selected for used.However, during preparation, a guidewire breakage was identified upon retraction of the packaging, and the device was unable to be used.The procedure was completed with another of same device.No patient complications were reported and the patient condition post procedure was stable.
 
Event or Problem Description
It was reported that guidewire break occurred.A 300cm pt2 moderate support guidewire was selected for used.However, during preparation, a guidewire breakage was identified upon retraction of the packaging, and the device was unable to be used.The procedure was completed with another of same device.No patient complications were reported and the patient condition post procedure was stable.
 
Additional Manufacturer Narrative
A2 - age at time of event: 18 years or older.The device was returned for analysis.Visual inspection revealed multiple kinks noted in the body of the guidewire.Microscopic inspection showed no signs of detachment in the ends of the guidewire.Dimensional testing showed the device length met specification.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PT
Common Device Name
WIRE, GUIDE, CATHETER
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
302 parkway, global park
la aurora - heredia
CS  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15621729
Report Number2124215-2022-41997
Device Sequence Number20419447
Product Code DQX
UDI-Device Identifier08714729471257
UDI-Public08714729471240
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K030617
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/15/2023
Device Model Number2783
Device Catalogue Number2783
Device Lot Number0028561428
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer12/20/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date12/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/15/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-