Model Number 694958 |
Device Problems
Fracture (1260); High impedance (1291); Inappropriate/Inadequate Shock/Stimulation (1574); Defective Device (2588)
|
Patient Problems
Emotional Changes (1831); Shock from Patient Lead(s) (3162)
|
Event Date 09/14/2006 |
Event Type
Injury
|
Manufacturer Narrative
|
Concomitant medical products: product id: d334drg icd, implanted: (b)(6) 2014.Product event summary: the proximal portion of the lead was returned, analyzed and no anomalies were found.Visual analysis of the lead indicated apparent explant damage.The initial reported event of lead fracture was previously submitted via a remedial action exemption (rae) summary report.The manufacturer has voluntarily discontinued this rae, so supplemental information is being submitted via a 30 day report.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
(b)(6) 2009: a lawsuit alleged that the patient has sustained and will continue to sustain severe physical injuries, including inappropriate therapies, severe emotional distress, and economic and consequential damages due to the 'defective' and fractured lead.(b)(6) 2022, it was further reported that the patient was inappropriately shocked by the right ventricular (rv) lead.The lead had high pacing impedance.The lead had a fracture.The lead was removed and a new lead was implanted.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|