• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED) Back to Search Results
Model Number KD-VC631Q-07203S
Medical Device Problem Codes Break (1069); Flare or Flash (2942)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/13/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device has not yet been received for evaluation.The cause of the reported issue is unknown at this time.Follow up is in progress regarding the return and additional information regarding the reported event.Supplemental report(s) will be submitted should any relevant new information is available and or received.Investigation is ongoing.This report will be supplemented accordingly following investigation.
 
Event or Problem Description
Device was reported to fail upon initial clinical use.Olympus area salesperson customer representative (rep.) reported the cutting wire of the kd-vc631q-07203s had broken and burn in half at the distal during a procedure.There was no patient harm, no user injury reported due to the event.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the approved final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, and since the device was not returned for evaluation, the definitive root cause of the broken and burnt cutting wire issue could not be determined.It is possible that the cutting wire was broken because the device may have been energized by coming into point contact with or being to tissue or the distal end of the endoscope.When the forceps elevator was raised, the cutting wire where the wire coating was torn came into contact with the distal end of the endoscope and may have caused an electrical discharge in the cutting wire, causing the cutting wire to become hot instantly and break.The instruction manual identifies the following verbiage, which may have prevented the phenomenon: ¿¿since the cutting wire is very thin, it may break off in the following cases: the distance between the papilla of vater and the cutting wire is very short, the output is too high or activated while the cutting wire touches metal parts of the endoscope, or the cutting wire is tightened too strong.When the cutting wire breaks off, its proximal end will be retracted toward the endoscope if the slider is pulled.If the slider is pushed, the cutting wire will be pushed out toward the papilla or move sideways.If the cutting wire breaks off, stop the output immediately and pull the slider completely to retract the broken cutting wire into the tube.Then withdraw the sphincterotome from the papilla.Otherwise, patient injury, such as perforations, bleeding, or lacerations within the biliary duct, and/or damage of the endoscope could result.¿be sure that the rear end of the cutting wire is extended from the distal end of the endoscope.In case the cutting wire contacts the forceps elevator, insufficient output or unintended tissue injury may occur.¿ do not activate output while the cutting wire touches the metal parts of the endoscope, or they are being close together.This could burn the tissue and/or damage the endoscope or the instrument.¿when inserting the instrument into the endoscope, be sure that the cutting wire is parallel to the tube.Otherwise, the metal part of the forceps elevator may contact the cutting wire and peel off the coating material.¿do not activate output while tissue is in contact with the torn or damaged coated portion of the distal end.If output is activated while tissue is contacting the torn or damaged coated portion due to insertion into or withdrawal from an endoscope, leakage current, decreased output, and/or thermal injury could result.¿if you feel the cutting is blunt, withdraw the device from the scope to examine if there is any peel off and tear at the coating portion.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE PRELOADED SPHINCTEROTOME V(DISTAL WIREGUIDED)
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer Contact
masaharu hirose
2-248-1 okkonoki
kuroishi-shi, aomori 036-0-357
JA   036-0357
426422891
MDR Report Key15621881
Report Number9614641-2022-00495
Device Sequence Number20429501
Product Code KNS
UDI-Device Identifier04953170371738
UDI-Public04953170371738
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K141991
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/13/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-VC631Q-07203S
Device Lot Number26V
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer11/24/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date12/13/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured06/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-