• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM TIB BASEPLATE - CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON PRIM TIB BASEPLATE - CEMENTED; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Model Number 5520-B-300
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Ambulation or Postural Difficulties (2544)
Date of Event 09/20/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Event or Problem Description
It was reported by the sales rep that surgeon notified about the tkr revision is needed.Update: reason for revision due to dislocation.
 
Additional Manufacturer Narrative
Reported event: an event regarding dislocation involving a triathlon baseplate was reported.The event was confirmed via clinician review of the provided medical records.Method & results: device evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: event description: it was reported by the sales rep that surgeon notified about the tkr.Revision is needed.Anatomical position: left knee.Product(s) involved: triathlon cr femur #3 left cemented, primary tibial baseplate cemented, cs insert x3 #3 x 11mm materials reviewed: undated/unlabelled: ap left knee x-ray, left lateral x-ray relevant clinical history and timelines: x-rays provided showed a lateral tka with an obvious posterior knee dislocation, complete disarticulation of the femur and tibia in the sagittal plane.There may be a quadriceps rupture based on a bony fragment.There is evidence of a mako procedure based on pin holes.It is unclear if the ap film is from the same event as in this plane there is no gross abnormality.No additional history or medical records were not provided.Date of implant: (b)(6) 2022 (explant (b)(6) 2022).Additional medical interventions noted: revision surgery (b)(6) 2022.Are any additional medical interventions involving stryker devices noted in clinical history: no.Conclusion of assessment: a knee dislocation occurred in an elderly patient within 4 1/2 months of the index operation.No other data were provided for review.Device or procedure related adverse event: triathlon cr cemented tka with cs polyethylene.Root cause and confirmation: a root cause cannot be ascertained without additional information.However, a knee dislocation may be confirmed.Device history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusion: no further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported by the sales rep that surgeon notified about the tkr revision is needed.Update: reason for revision due to dislocation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON PRIM TIB BASEPLATE - CEMENTED
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-LIMERICK
raheen business park
limerick NA
EI   NA
Manufacturer Contact
diana rogers
210 centennial avenue
centennial park, elstree
borehamwood, NJ WD6 3-SJ
UK   WD6 3SJ
2018315000
MDR Report Key15622121
Report Number0002249697-2022-01506
Device Sequence Number6499298
Product Code MBH
UDI-Device Identifier07613327050318
UDI-Public07613327050318
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model Number5520-B-300
Device Catalogue Number5520-B-300
Device Lot NumberNHE2S
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer12/09/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date01/05/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured07/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age87 YR
Patient SexFemale
-
-