EXACTECH, INC. LOGIC TIBIA IMPLANT PSC INSERT, SZ 2.5, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED
|
Back to Search Results |
|
| Model Number |
02-012-44-2511 |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Aseptic loosening (4883)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
As reported, the patient had an initial right tka on (b)(6) 2018.The patient was revised due to wear and loosening and the tibial insert and baseplate were replaced.The patient was last known to be in stable condition following the event.No other patient information/medical history provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Pending investigation.Concomitant medical devices: 02-012-41-2525 - logic tibia traptray cem sz 2.5f/2.5t 4509768, 02-020-11-0325 - truliant ps cem fem ps cem right sz 2.5 5679357, 200-02-35 - three peg patella 35mm 5393556.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H3: the revision reported was likely the result of prosthesis wear, osteolysis, and an insufficient bond between the tibial component and the bone, which led to aseptic (non-infected) tibial loosening.Possible causes for polyethylene wear include malalignment between the implants, high contact stresses during knee flexion, third body wear, patient-related conditions, instability, or any combination of these possibilities.Additionally, a contributing factor to the wear may have been the result of being packaged in a non-conforming vacuum bag for more than five years.The extent and root cause of the prosthesis wear, osteolysis, and tibial loosening could not be determined as the devices were not returned for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, d8, f6, f9 (remove erroneous dates from initial report) h6: clinical code, component code, investigation findings, investigation conclusions, h10.H10: multiple mdr reports were filed for this event, please see associated report: 1038671-2026-00293.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|
Event or Problem Description
|
|
Report #1 of 2 for this event.As reported, the patient had an initial right total knee arthroplasty.Subsequently, three (3) years, ten (10) months later, the patient underwent a right knee revision of the tibial components.The patient was last known to be in stable condition following the event.Images of the explants were provided and pre-operative radiographs were provided.No additional clinical information was provided.
|
| |
|