| Model Number |
80619 |
| Medical Device Problem Codes |
Fracture (1260); Obstruction of Flow (2423); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/29/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A direxion hi-flo was selected for a uterine fibroid embolization procedure.While injecting beadblock, the catheter clogged several times.Intermittent flushing was performed to resolve the clogging.During removal from the patient, the physician noticed that the outer shaft was broken and detached from the inner shaft.There had not been a lot of torquing of the catheter by the interventional radiologist.There were no patient complications, and the procedure was completed successfully.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A direxion hi-flo was selected for a uterine fibroid embolization procedure.While injecting beadblock, the catheter clogged several times.Intermittent flushing was performed to resolve the clogging.During removal from the patient, the physician noticed that the outer shaft was broken and detached from the inner shaft.There had not been a lot of torquing of the catheter by the interventional radiologist.There were no patient complications, and the procedure was completed successfully.It was further reported that 3 syringes of beadblock mixed with contract were delivered using the microcatheter.There was no resistance or friction experienced during removal of the catheter.The anatomy was mildly tortuous.The patient status was okay post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the direxion microcatheter was returned to boston scientific for analysis.The device shaft was visually and microscopically analyzed for any damage.The devices shaft showed a fracture located 40cm from the hub.The device was not completely separated the inner liner was still attached.Inspection of the remainder of the device, apart from the observed damage revealed no other damage or irregularities.
|
| |
|
Event or Problem Description
|
|
It was reported that a shaft break occurred.A direxion hi-flo was selected for a uterine fibroid embolization procedure.While injecting beadblock, the catheter clogged several times.Intermittent flushing was performed to resolve the clogging.During removal from the patient, the physician noticed that the outer shaft was broken and detached from the inner shaft.There had not been a lot of torquing of the catheter by the interventional radiologist.There were no patient complications, and the procedure was completed successfully.It was further reported that 3 syringes of beadblock mixed with contract were delivered using the microcatheter.There was no resistance or friction experienced during removal of the catheter.The anatomy was mildly tortuous.The patient status was okay post procedure.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: describe event or problem: additional information.
|
| |
|