ETHICON ENDO-SURGERY, LLC. 12MM XCEL STABILITY SLEEVE; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Model Number CB12LT
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/28/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Batch #: unk.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: "were there any patient consequences? if yes, please describe." the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample determined that cb12lt device was returned inside its package unopened and with the universal seal unlatch from the sleeve.In an attempt to replicate the reported incident, the device was functionally tested to detect any reducer attachment removal issues.Upon evaluation of the device, the universal seal latch onto the sleeve assembly.In addition the obturator latch onto the universal seal as well.The device was fully functional according to the manufacturing requirements.Upon visual inspection, it was observed that the pouch and tyvek from the packaging were damaged; it was noted to have a hole in the tyvek, the hole was noted to be from the inside out.Due to the damages found on the packaging, a possible cause for these conditions is due to improper handling during transit or storage; it appears that the package hit a pointy surface and this caused the reported event.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
 
Event or Problem Description
It was reported that during a hernia repair, damage was noticed on the top seal; part of the device had broken off into the sterile packaging.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
12MM XCEL STABILITY SLEEVE
Common Device Name
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key15624489
Report Number3005075853-2022-06965
Device Sequence Number6470115
Product Code GCJ
UDI-Device Identifier10705036001331
UDI-Public10705036001331
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K032676
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2026
Device Model NumberCB12LT
Device Catalogue NumberCB12LT
Device Lot Number489A44
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/22/2022
Initial Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/08/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-