• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC CATH PKGD: WEDGE 6 FR 110 CM; CATHETER, FLOW DIRECTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC CATH PKGD: WEDGE 6 FR 110 CM; CATHETER, FLOW DIRECTED Back to Search Results
Model Number IPN923527
Medical Device Problem Code Failure to Deflate (4060)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 08/31/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Qn# (b)(4).Medwatch report key/report number: (b)(4).
 
Event or Problem Description
Complaint from medwatch report.The report states that during a "right heart catheterization, when the arrow- balloon wedge pressure catheter was used, only part of the balloon would inflate, then it would not deflate properly.No known harm to patient but did delay case".Additional information states that a new catheter was inserted without issue.No report of patient harm or injury.Associated mdr 3010532612-2022-00427.
 
Event or Problem Description
Complaint from medwatch report.The report states that during a "right heart catheterization, when the arrow- balloon wedge pressure catheter was used, only part of the balloon would inflate, then it would not deflate properly.No known harm to patient but did delay case".Additional information states that a new catheter was inserted without issue.No report of patient harm or injury.Associated mdr 3010532612-2022-00427.
 
Additional Manufacturer Narrative
Qn#(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH PKGD: WEDGE 6 FR 110 CM
Common Device Name
CATHETER, FLOW DIRECTED
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15624892
Report Number3010532612-2022-00420
Device Sequence Number6345592
Product Code DYG
UDI-Device Identifier10801902206746
UDI-Public10801902206746
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K892530
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/23/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberIPN923527
Device Catalogue NumberAI-07126
Device Lot Number16F22E0079
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer11/11/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date11/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/16/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age74 YR
Patient SexMale
Patient Weight67 KG
-
-