• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR ANCHOR W/ ORTHOCORD AND NEEDLES; SOFT-TISSUE ANCHOR, BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR ANCHOR W/ ORTHOCORD AND NEEDLES; SOFT-TISSUE ANCHOR, BIOABSORBABLE Back to Search Results
Model Number 222299
Medical Device Problem Codes Break (1069); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/20/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi:(b)(4).Reporter is a j&j sales representative.Investigation summary: the complaint device was received and evaluated.Upon visual inspection, it could be observed that the inserter shaft is not deformed or bent, the inserter tip is in good condition, no structural anomalies were found.The anchor is broken.Rests of biological matter can be observed on the anchor.A manufacturing record evaluation was performed for the finished device 8l09690 number, and no non-conformances were identified.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.A possible root cause for the broken anchor can be attributed to procedural variables, such handling of the device or product interaction during procedure; axial misalignment or bending force was applied, as per ifu; axial misalignment or levering with the anchor upon insertion may result in anchor fracture.Do not apply a bending force to the inserter.This can damage the anchor or inserter tip.At this point in time, no corrective action is required, and no further action is warranted.However, in depuy synthes mitek, additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Event or Problem Description
It was reported by the sales rep in south korea that during a hand procedure on (b)(6) 2022, it was observed that the 5.5 healix advance br anchor w/ orthocord and needles device had an unspecified malfunction.During in-house engineering evaluation, it was determined that the anchor on the device was broken.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE BR ANCHOR W/ ORTHOCORD AND NEEDLES
Common Device Name
SOFT-TISSUE ANCHOR, BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15625013
Report Number1221934-2022-03140
Device Sequence Number6345602
Product Code MAI
UDI-Device Identifier10886705021352
UDI-Public10886705021352
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K120078
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2024
Device Model Number222299
Device Catalogue Number222299
Device Lot Number8L09690
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/26/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/02/2022
Initial Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-