Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event site postal code: (b)(6).Event site telephone number: (b)(6).The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id #: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during intra-aortic balloon (iab) therapy, a leak occurred.A new iab was inserted to continue therapy.There was no patient harm or adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Reference complaint (b)(4).
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|
Additional Manufacturer Narrative
|
|
The iab was returned with the membrane completely unfolded.No blood was visible on the iab catheter.An underwater leak test of the balloon, catheter, y-fitting and extracorporeal tubing was performed and no leaks were detected.The reported event cannot be confirmed by the evaluation.A lot history record review was completed for the reported product.No nonconformances were found that are considered to be related to the event.Reference complaint: (b)(4).
|
| |
|