| Model Number |
23AGFN-756 |
| Medical Device Problem Code |
Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2022, a 23mm regent valve was implanted in the aortic.The leaflet was opening and closing properly when the device was implanted.On an unknown date in 2022, two weeks after surgery, one leaflet became immobile.The patient was controlled at inr 1.96 or above.Thrombosis was not confirmed by imaging, but thrombolytic therapy was provided with no improvement observed.Therefore, on (b)(6) 2022, the device was explanted from the patient and a non-abbott device was implanted as a replacement.The device was damaged during extraction.The leaflet may be damaged when the device was washed.The patient's condition is currently stable.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of one leaflet becoming immobile was reported.The investigation found that both the leaflets were intact and fully opening and closing.Chipping was noted on the leaflets.No damage was seen on the orifice.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and that the product met all specifications.The cause of the reported incident could not be conclusively determined.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
| |
|
Event or Problem Description
|
|
N/a.
|
| |
|