• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM KOREA; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO SURGICAL SYSTEM KOREA; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB00079
Device Problem Circuit Failure (1089)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2022
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event Description
It was reported that, during a navio assisted surgery, the siu has pfs error during booting.There was no red light in siu and the fan is working.All cables appear to be tightly connected and siu power is on.The procedure was completed manually, with a delay of less than 30 min.Patient was not harmed.
 
Manufacturer Narrative
The navio surgical system korea, part number rob00079, serial number (b)(6), used for treatment was not returned for evaluation.A relationship between the reported event and the device could not be established.A complaint history review for similar reported/confirmed complaints has identified prior events.A review of manufacturing records indicate the device met all specifications upon release into distribution.The navio surgical technique guide for knee arthroplasty (500197 rev c) provides guidelines for recovering to a fully manual procedure in the ¿recovery procedure guidelines¿.The failure mode and associated risk have been anticipated within the risk file and the documented risk level is still within the product risk profile.Although the reported problem was not confirmed through a visual or functional evaluation, factors contributing to the reported symptom may have been associated with corrupt cpu bios settings.Based on the investigation, the need for a corrective action is not recommended or required at this time.Should any additional information be received the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO SURGICAL SYSTEM KOREA
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15625640
MDR Text Key301927254
Report Number3010266064-2022-00629
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556725597
UDI-Public885556725597
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
K201022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 11/17/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberROB00079
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 09/22/2022
Initial Date FDA Received10/18/2022
Supplement Dates Manufacturer Received11/13/2022
Supplement Dates FDA Received11/17/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
ASSY, PRGMMED, SIU, COMP CART, NAVIO.
Patient Outcome(s) Required Intervention;
-
-