Catalog Number 955365
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/23/2022
Event Type  malfunction  
Manufacturer Narrative
Initial reporter phone no.: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that one unit of theranova 400, had an external saline leak where the arterial line was connected.The event occurred before patient use.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
H10: the actual device and pictures were returned for evaluation.The visual inspection of the actual sample by naked eyes showed the product as on the provided photo dry and the blood protection caps was attached on the dialysate port.A torn piece of silicone was found on one blood connection, on the inner cone.The most likely cause of the reported failure is the torn piece of silicone of the arterial line connector, which probably tore off due to over-tightening of the connector and caused the external leakage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERANOVA
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - HECHINGEN BADEN DIALYZER
holger - crafoord - strasse 26
hechingen D-723 79
GM   D-72379
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 
2242702068
MDR Report Key15625769
MDR Text Key302927992
Report Number9611369-2022-00195
Device Sequence Number1
Product Code KDI
UDI-Device Identifier05413765567841
UDI-Public(01)05413765567841
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 11/09/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/18/2022
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955365
Device Lot Number1-6506-H-01
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/28/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
NA.
-
-