Model Number 71453222
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that, after a tka surgery was performed on 2012, one (1) lgn ps high flex xlpe sz 5-6 11mm broke.This adverse event was treated by a revision surgery on (b)(6) 2022, in which the lgn ps high flex xlpe sz 5-6 11mm was exchanged.No delay reported.Patient's current health status is unknown.
 
Additional Manufacturer Narrative
H3, h6: the associated device was returned and evaluated.The visual inspection revealed the post of the device has fractured off.A lab analysis performed on the device revealed that the articulating and non-articulating surfaces of the insert are worn with some discoloration.The post was fractured at the base of the insert and exhibited signs of wear and deformation.There were no observations of material or manufacturing deviations in the course of this investigation.The clinical/medical investigation concluded that, the provided undated x-ray and two photos were reviewed but do not aid in the clinical investigation of the root cause of the reported fracture of post.As of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.The patient's current condition is unknown, and the patient impact beyond the surgery and recovery, could not be determined based on the limited information provided.No further medical assessment could be rendered at this time.Should any additional clinical information be provided, this complaint will be re-evaluated a review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar a event for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that has been identified as a in warnings and precautions that the implant can break or become damaged as a result of strenuous activity or trauma.The patient should be warned that the device does not replace normal healthy bone.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.Verifying the device configuration according to the print is within the steps of the final inspection.Also, the quality and manufacture of 7.5 mrad cross-linked uhmwpe bar should be controlled according to the material specification.A review made by the quality engineering team revealed that the root cause of the fracture of the post could not be determined based on the limited information provided.Also there were no observations of material or manufacturing deviations in the course of this investigation.At this time, we have no reason to suspect that the product failed to meet any specifications at the time of manufacture.With the results of this investigation the root cause of this event could not be determined.Factors that could contribute to the reported event include traumatic injury or abnormal loading of limb.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LGN PS HIGH FLEX XLPE SZ 5-6 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key15626136
Report Number1020279-2022-04444
Device Sequence Number5978746
Product Code JWH
UDI-Device Identifier00885556035351
UDI-Public00885556035351
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K071071
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/23/2022
Device Model Number71453222
Device Catalogue Number71453222
Device Lot Number12KM18860
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer03/07/2023
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date03/13/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-