Model Number MODEL 100
Medical Device Problem Code Visual Prompts will not Clear (2281)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported a complaint that "the autopulse platform (sn (b)(4)displayed "system error, out of service, revert to manual cpr" was confirmed during the functional testing.The root cause for the reported complaint was due to a damaged coil on the processor board, likely due to normal wear and tear.The platform was manufactured in jan.2008, and is over 14 years old, past its expected service life of 5 years.Visual inspection of the returned platform was performed, and no physical damage was observed.The power distribution board revision was up to date and the brake gap was within specification.The archive data review could not be completed because the archive data could not be retrieved/read.Due to this, the specific system error message could not be determined.The autopulse platform failed initial functional testing due to the "system error, out of service, revert to manual cpr" error message displayed upon powering on, thus confirming the reported complaint.The damaged processor board was replaced to address the system error.Following service, the autopulse platform passed the run-in test without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint, and there were no similar complaints reported for autopulse platform with sn (b)(4).
 
Event or Problem Description
During shift check, the customer reported that the autopulse platform (sn (b)(4)displayed "system error, out of service, revert to manual cpr".No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Common Device Name
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLLCIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key15626277
Report Number3010617000-2022-01896
Device Sequence Number6225085
Product Code DRM
UDI-Device Identifier00849111000079
UDI-Public00849111000079
Combination Product (Y/N)N
Initial Reporter CountryNL
PMA/510(K) Number
K112998
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0700-16
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/20/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Initial Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/18/2008
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-