• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO PHILIPPINES CORPORATION MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO PHILIPPINES CORPORATION MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE; SYRINGE, ANTISTICK Back to Search Results
Model Number N/A
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/12/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implanted date: device was not implanted.Explanted date: device was not explanted.Establishment name: (b)(6).Health professional: requested, unknown.Reporter occupation: administrative coordinator.The root cause of the complaint could not be identified to be related to our production or process.The actual sample was not available for our evaluation.Retention samples were visually confirmed free from defects that will affect the activation of the safety sheath and passed the evaluation for sheath activation and deactivation.Also, no irregularity was encountered during the simulation of manual sheath activation that may lead to the complaint.We have a series of visual in-process inspections to detect an abnormality on the sheath that may lead to a problem during sheath activation.The molding condition of the components critical to the safety activation of the product is routinely checked to assure that no defects will be encountered that will lead to sheath activation problems.Similarly, the assembly status of the safety needle such as sheath collar fitting and damaged parts that will affect product function is being confirmed.We have received four (4) complaints covering fy20 to fy22 (september 28, 2022), related to this issue where most of the problems are related to usage particularly due to improper activation of the device (not activating on the hard and flat surface in a quick and firm motion).Based on the investigation of individual complaints there was no attributable cause noted related to our product and production process.Lot history file revealed no related nonconformity or irregularity that will lead to the complaint.Prior to shipment, qc conducts outgoing visual, sensory, and functional inspections to assure lots are of good quality.Therefore, we advise to follow the instructions for use (ifu) for the proper usage of the sg2 needle indicated on the unit box in which warnings to prevent needle stick, cautions, and precautions are also included.(b)(4).
 
Event or Problem Description
The user facility reported that the involved needle is very dangerous and hard to put the safety on.An associate almost incurred a needle stick.Additional information was received on 22 sept 2022: it was reported that the needle was new for the facility.Post- successful patient flu vaccine.The healthcare provider (hcp) had a near miss with the safety activation.The hcp was activating it on a hard surface, when the needle bent keeping it from activating, and did not get stuck.There was no heath impact to the actual patient or the hcp.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCKESSON PREVENT ® HT SAFETY HYPODERMIC NEEDLE
Common Device Name
SYRINGE, ANTISTICK
Manufacturer (Section D)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan,
RP 
Manufacturer (Section G)
TERUMO PHILIPPINES CORPORATION
124 east main avenue
laguna technopark
binan, laguna,,
RP  
Manufacturer Contact
gina digioia
265 davidson ave
suite 320
somerset, NJ 08873
6402040886
MDR Report Key15626302
Report Number3003902955-2022-00046
Device Sequence Number14947036
Product Code MEG
UDI-Device Identifier30612479178855
UDI-Public30612479178855
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K051865
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/18/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2026
Device Model NumberN/A
Device Catalogue Number102-N251S
Device Lot Number210807B
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/07/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-