SMITH & NEPHEW, INC. GNS II NON-POR TIB SZ 7 RIGHT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
|
Back to Search Results |
|
| Model Number |
71420190 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Codes |
Pain (1994); Arthralgia (2355); Inadequate Osseointegration (2646); Joint Laxity (4526)
|
| Date of Event |
09/22/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tka surgery had been performed on (b)(6) 2019, the patient experienced pain and the bia was loose.This adverse event was solved by a revision surgery on (b)(6) 2022.Current health status of patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.The clinical medical team reviewed the single undated/unidentified ap x-ray image provided appears to show a right cemented tka with a reduced lateral tibia fracture inferior to the baseplate, however, no comparison images were provided.The clinical/medical investigation concluded that, based on the limited documentation provided, definitive clinical factors could not be concluded to have contributed to the reported event.The patient impact beyond the reported pain, ¿tibia loose¿, the noted lateral tibial fracture on x-ray and the revision could not be determined.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that pain and looseness of components have been identified in possible adverse effects section.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include abnormal motion over time, bone degeneration, size selected, inadequate integration between the cement and bone/implant, osteolysis, patient condition, postoperative care and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a tka surgery had been performed on (b)(6) 2019, the patient experienced pain and the tibia was loose.This adverse event was solved by a revision surgery on (b)(6) 2022.Current health status of patient is unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Section h3, h6: the device was not returned for evaluation.The clinical medical team reviewed the x-ray image provided appears to show a right cemented tka with a reduced lateral tibia fracture inferior to the baseplate, however, no comparison images were provided.The clinical/medical investigation concluded that, based on the limited documentation provided, definitive clinical factors could not be concluded to have contributed to the reported event.The patient impact beyond the reported pain, ¿tibia loose¿, the noted lateral tibial fracture on x-ray and the revision could not be determined.No further medical assessment can be rendered at this time.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the instructions for use documents for knee systems revealed that pain and looseness of components have been identified in possible adverse effects section.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Factors that could contribute to the reported event include abnormal motion over time, bone degeneration, size selected, inadequate integration between the cement and bone/implant, osteolysis, patient condition, postoperative care and/or traumatic injury.The contribution of the device to the reported event could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.B5: describe event or problem.
|
| |
|