It was reported that during a tka, when impacting the tibial tray, shards of plastic were coming off the gii tibial base impactor and falling in the wound.All broken pieces removed with forceps.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.
|
|
Section h3, h6: the associated device was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.Several small fractures occurred on the plastic head of the device.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.The clinical/medical evaluation concluded that as of the date of this medical investigation, the requested clinical documentation has not been provided; therefore, there were no clinical factors found which would have contributed to the event.According to the report, all the broken pieces were removed from inside of the patient with forceps, and the procedure was completed without any delay using the same device.Since there has been no harm alleged because of this event, no further clinical/medical assessment is warranted at this time.Should any additional relevant medical information be provided, this case would be re-assessed.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.Assessment of historical escalated cases concluded that there are no prior actions related to this device and failure mode.As the device broke and it can no longer fit its purpose, the contribution of the device to the reported event could be corroborated.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
|