MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE Back to Search Results
Model Number 0103L24
Medical Device Problem Code Deflation Problem (1149)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/06/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient had a foley catheter with 30 cc balloon inserted due to straight catheter unsuccessful.On (b)(6) 2022, the patient was pulling at foley and causing trauma and developed sang discharge in foley.When attempted to remove foley, it was unable to remove due to bulb not deflating.Confirmed on ultrasound that bulb was full with fluid.Attempted client repositioning and multiple attempts to withdraw fluid.Surgical consult was recommended to have an ultrasound guided procedure completed.On (b)(6) 2022, the patient was left to facility 2 for surgical consult.Bulb was deflated and new foley inserted.Removed when back in facility.Incident was thought to have occurred due to faulty equipment and was able to remove about 2.5 ml or less of fluid, unable to remove more.Bulb was not tested/inflated prior to insertion.This was not done due to recent changes to best practice for foley catheter insertion.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported that the patient had a foley catheter with 30 cc balloon inserted due to straight catheter unsuccessful.On (b)(6) 2022, the patient was pulling at foley and causing trauma and developed sang discharge in foley.When attempted to remove foley, it was unable to remove due to bulb not deflating.Confirmed on ultrasound that bulb was full with fluid.Attempted client repositioning and multiple attempts to withdraw fluid.Surgical consult was recommended to have an ultrasound guided procedure completed.On (b)(6) 2022, the patient was left to facility 2 for surgical consult.Bulb was deflated and new foley inserted.Removed when back in facility.Incident was thought to have occurred due to faulty equipment and was able to remove about 2.5 ml or less of fluid, unable to remove more.Bulb was not tested/inflated prior to insertion.This was not done due to recent changes to best practice for foley catheter insertion.
 
Additional Manufacturer Narrative
The reported event was inconclusive.No sample was returned for evaluation.The potential root cause for this failure could be "lumen collapses under the pressure of the balloon, pinched lumen;".It was unknown whether the device had met specifications.The product was used for treatment purposes.It was unknown whether the product had caused the reported failure.The dhr review could not be performed without a lot number.The instructions for use were found adequate and state the following: "to deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRICATH® FOLEY CATHETER TIEMANN MODEL COUDE
Common Device Name
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key15626905
Report Number1018233-2022-07956
Device Sequence Number6184742
Product Code KOD
UDI-Device Identifier00801741018220
UDI-Public(01)00801741018220
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K922431
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,User Facility
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0103L24
Device Catalogue Number0103L24
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/06/2022
Supplement Date Received by Manufacturer02/22/2023
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date02/27/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-