• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL LLC AUTOCAT2 WAVE JAPANESE; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL LLC AUTOCAT2 WAVE JAPANESE; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN000325
Medical Device Problem Code Pumping Stopped (1503)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/02/2022
Type of Reportable Event Malfunction
Event or Problem Description
The report states "after 17:00 on (b)(6) 2022, an urgent patient came and an iab was inserted.About 23:00, while driving iabp, the screen suddenly froze, and pumping stopped.The user pushed the button, but it did not work at all, so the user turned off the power once and restarted, then the drive resumed.About 0:30 on (b)(6) 2022, about 1 hour after restart, pumping stopped again.The screen did not freeze but pushing the button did not work.Therefore, iabp was replaced with another one in the facility".No report of patient harm or injury.
 
Additional Manufacturer Narrative
(b)(4).N/a.Other remarks: n/a.Corrected data: n/a.
 
Additional Manufacturer Narrative
(b)(4).Complaint verification testing could not be performed as no sample was returned for analysis.Additionally no recorder strip was returned for investigation.A device history record review was performed, and no relevant findings were identified.Without the device to evaluate, the complaint could not be confirmed, and the probable cause could not be determined from the available information.Teleflex will continue to monitor and trend for reports of this nature.
 
Event or Problem Description
The report states "after 17:00 on (b)(6) 2022, an urgent patient came and an iab was inserted.About 23:00, while driving iabp, the screen suddenly froze, and pumping stopped.The user pushed the button, but it did not work at all, so the user turned off the power once and restarted, then the drive resumed.About 0:30 on 03 oct 2022, about 1 hour after restart, pumping stopped again.The screen did not freeze but pushing the button did not work.Therefore, iabp was replaced with another one in the facility".No report of patient harm or injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOCAT2 WAVE JAPANESE
Common Device Name
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
bryanna connelly
3015 carrington mill blvd
morrisville 27560
MDR Report Key15627009
Report Number3010532612-2022-00436
Device Sequence Number20457621
Product Code DSP
UDI-Device Identifier30801902043420
UDI-Public30801902043420
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K060309
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/04/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberIPN000325
Device Catalogue NumberIAP-0500J
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer11/23/2022
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date11/23/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
N/A.; N/A.
Patient SexUnknown
-
-