| Model Number |
MS1-5590S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Metal Related Pathology (4530)
|
| Date of Event |
10/11/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
Information was received that there were signs of metallosis and slight pistoning.The rod has been explanted.
|
| |
|
Event or Problem Description
|
|
No additional information has been provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional data: b5, h6, h10.The unit was not returned to nuvasive for investigation.This investigation has been conducted by evaluation of provided x-ray image.It was noted that the rod has been distracted beyond the mid-point.There were no obvious device-related abnormalities noted.No objective evidence was found to indicate the presence of metallosis or locking pin failure (slight pistoning).The reported event is unable to be confirmed.The root cause is unable to be determined.Review of the device history records indicated that the unit met all quality specifications and testing prior to product departure.
|
| |
|