| Model Number |
IPN000085 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The report states that "the balloon inflated asymmetrically during use.Therefore, the catheter was replaced with another kit of the same lot, inserted at the same insertion site to complete the procedure.The user flushed the catheter but did not perform the inflation test before use".No delay in therapy.No report of patient harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Qn# (b)(4).
|
| |
|
Event or Problem Description
|
|
The report states that "the balloon inflated asymmetrically during use.Therefore, the catheter was replaced with another kit of the same lot, inserted at the same insertion site to complete the procedure.The user flushed the catheter but did not perform the inflation test before use".No delay in therapy.No report of patient harm or injury.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).The reported complaint that the "balloon inflated asymmetrically during use" is not able to be confirmed.The product was not returned for investigation.Based on a review of the device history record (dhr), the product met specification upon release.The root cause of the complaint is undetermined.The reported complaint will be monitored for any developing trends.Additionally, according to the event details, the user "did not perform the inflation test before use".As a result, an in-service has been requested to review the instructions for use (ifu) with the customer.Other remarks: n/a.Corrected data: n/a.
|
| |
|