| Model Number |
110-180 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
07/25/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported right side "rupture." device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Additional, changed, and/or corrected data: a2, b3, b5, b6, d1, d2, d4, d6a d6b, g4, h4, h6.
|
| |
|
Event or Problem Description
|
|
The device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: was observed, broken device on the posterior side assessed as unidentified (tear) opening (shell thickness within specification).Additional observations: missing shell assessed as inconclusive.Wear abrasion is observed.No further actions are required as the device was implanted.
|
| |
|
Event or Problem Description
|
|
Patient reported right side "rupture." the device has been explanted and replaced.
|
| |
|