| Model Number |
40458 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Thrombosis/Thrombus (4440)
|
| Date of Event |
09/01/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr.: the complaint device was not returned for analysis.The costumer attached a picture of the device and it is possible to observed that de distal tip was detached; consequently, confirming the reported distal tip "detachment of device or device component" complaint event.
|
| |
|
Event or Problem Description
|
|
It was reported that distal tip detachment and stent thrombosis occurred.The target lesion was located in the leg.A 300cm straight tip v-14 control wire guidewire was advanced for treatment.However, during the procedure, the distal tip of the wire was broken inside the vessel and thrombosis was noted.The detached portion was removed using a snare and the procedure was completed using alternate method or device.No further patient complications reported and the patient was fully recovered.
|
| |
|
Event or Problem Description
|
|
It was reported that distal tip detachment and stent thrombosis occurred.The target lesion was located in the leg.A 300cm straight tip v-14 control wire guidewire was advanced for treatment.However, during the procedure, the distal tip of the wire was broken inside the vessel and thrombosis was noted.The detached portion was removed using a snare and the procedure was completed using alternate method or device.No further patient complications reported and the patient was fully recovered.
|
| |
|
Additional Manufacturer Narrative
|
|
B3 date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr: the complaint device was not returned for analysis.The costumer attached a picture of the device, and it was observed that the distal tip was detached.H3 other text : photo provided.
|
| |
|