EXACTECH, INC. PS TIBIAL INSERTS SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
|
Back to Search Results |
|
| Model Number |
PS TIBIAL INSERTS SZ 4, 11MM |
| Medical Device Problem Code |
Naturally Worn (2988)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Insufficient Information (4580)
|
| Date of Event |
05/28/2022
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a patient, initial knee implanted on (b)(6) 2016, underwent a revision procedure on (b)(6) 2022, approximately 5 years 11 months post the initial implant procedure for poly wear.No further information reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information, including the product investigation, will be submitted within 30 days of receipt.
|
| |
|
Additional Manufacturer Narrative
|
|
After review of additional information received the following sections, have been updated accordingly: h6: the revision reported was likely the result of wear of the tibial insert.However, this cannot be confirmed because the component was not returned for evaluation and no x-rays or photographs were provided.Additionally, a contributing factor to the wear may have been the result of the insert being packaged in a non-conforming vacuum bag for more than five years.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient, had an initial knee implanted and then underwent a revision procedure, approximately 5 years 11 months post the initial implant procedure for poly wear.No further information reported.
|
| |
|
Additional Manufacturer Narrative
|
|
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b5, g1, g2, h6: clinical code, component code, type of investigation, investigation findings, and investigation conclusions.The revision reported was likely the result of wear of the tibial insert.However, this cannot be confirmed because the component was not returned for evaluation and no x-rays or photographs were provided.Additionally, a contributing factor to the wear may have been the result of the insert being packaged in a non-conforming vacuum bag for more than five years.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
|
| |
|