MAUDE Adverse Event Report: EXACTECH, INC. PS TIBIAL INSERTS SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. PS TIBIAL INSERTS SZ 4, 11MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, Back to Search Results
Model Number PS TIBIAL INSERTS SZ 4, 11MM
Medical Device Problem Code Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 05/28/2022
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a patient, initial knee implanted on (b)(6) 2016, underwent a revision procedure on (b)(6) 2022, approximately 5 years 11 months post the initial implant procedure for poly wear.No further information reported.
 
Additional Manufacturer Narrative
Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Additional Manufacturer Narrative
After review of additional information received the following sections, have been updated accordingly: h6: the revision reported was likely the result of wear of the tibial insert.However, this cannot be confirmed because the component was not returned for evaluation and no x-rays or photographs were provided.Additionally, a contributing factor to the wear may have been the result of the insert being packaged in a non-conforming vacuum bag for more than five years.
 
Event or Problem Description
It was reported that a patient, had an initial knee implanted and then underwent a revision procedure, approximately 5 years 11 months post the initial implant procedure for poly wear.No further information reported.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: b1, b2, b5, g1, g2, h6: clinical code, component code, type of investigation, investigation findings, and investigation conclusions.The revision reported was likely the result of wear of the tibial insert.However, this cannot be confirmed because the component was not returned for evaluation and no x-rays or photographs were provided.Additionally, a contributing factor to the wear may have been the result of the insert being packaged in a non-conforming vacuum bag for more than five years.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PS TIBIAL INSERTS SZ 4, 11MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED,
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
geoff gannon
3523771140
MDR Report Key15628877
Report Number1038671-2022-01325
Device Sequence Number2749272
Product Code JWH
UDI-Device Identifier10885862048394
UDI-Public10885862048394
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K933610
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2022
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/25/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/18/2018
Device Model NumberPS TIBIAL INSERTS SZ 4, 11MM
Device Catalogue Number204-24-11
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer02/21/2023
02/19/2026
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date03/13/2023
02/25/2026
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/19/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-0019-2022
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK.
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexUnknown
-
-