SYNTHES GMBH SOCKET WRENCH FOR VEPTR NUT Back to Search Results
Model Number 03.641.004
Medical Device Problem Code Calibration Problem (2890)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that on unknown date, during evaluation, the repair technician found that the socket wrench for veptr nut failed calibration.The issue was found during testing at service and repair.No additional information available.This report is for one (1) socket wrench for veptr nut.This is report 1 of 1 for complaint (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Reporter is a j&j employee.A product investigation was conducted.Service and repair evaluation: it was reported that during evaluation, the repair technician found that the device 03.641.004, socket wrench for veptr nut failed calibration.The issue was found during testing at service and repair.The repair technician reported that the device failed low for torque test during evaluation.The device minimum peak was at 4.864 nm which is lower than the accepted torque range of 5 nm- 6.4 nm.Torque test low is the reason for repair.The cause of the issue is torque test failed low.The item will be forwarded to customer quality.The evaluation was confirmed.A device history record (dhr) review was conducted: part # 03.641.004.Synthes lot # 9887049-11.Supplier lot # 9887049-11.Release to warehouse date: 07 jan 2016.Supplier: avalign technologies-nemcomed no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOCKET WRENCH FOR VEPTR NUT
Common Device Name
WRENCH
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kate karberg
1302 wrights lane east
west chester, PA 19380
8472871282
MDR Report Key15629054
Report Number8030965-2022-08388
Device Sequence Number5983113
Product Code HXC
UDI-Device Identifier10705034720128
UDI-Public(01)10705034720128
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Other
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.641.004
Device Catalogue Number03.641.004
Device Lot Number9887049-11
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/27/2022
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/28/2022
Initial Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/07/2016
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient SexUnknown
-
-