• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LUMENIS LTD LUMENIS SLIMLINE 200 D/F/L; POWERED LASER SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LUMENIS LTD LUMENIS SLIMLINE 200 D/F/L; POWERED LASER SURGICAL INSTRUMENT Back to Search Results
Model Number 1911-07
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/20/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during a procedure, when connecting the fiber to the console, the fiber broke and detached from the rubber connector that inserts into the laser console.Another fiber was used in order to complete the procedure.There were no patient complications reported.This report corresponds to the third of four fibers.
 
Event or Problem Description
It was reported that, during a procedure, when connecting the fiber to the console, the fiber broke and detached from the rubber connector that inserts into the laser console.Another fiber was used in order to complete the procedure.There were no patient complications reported.This report corresponds to the third of four fibers.
 
Additional Manufacturer Narrative
The product was not returned; therefore, no physical analysis of the product could be performed.Review of a photo of the device submitted showed that the fiber connector was detached from the fiber, confirming the reported clinical observations.Review of the device history record found no deviations that could have contributed to the reported event.Based on the available information, it is likely that a misaligned aiming beam caused the laser energy to heat an incorrect target ultimately leading to the detachment.Due to this, the most probable cause of cause traced to component failure has been applied to this complaint.
 
Event or Problem Description
It was reported that, during a procedure, when connecting the fiber to the console, the fiber broke and detached from the rubber connector that inserts into the laser console.Another fiber was used in order to complete the procedure.There were no patient complications reported.This report corresponds to the third of four fibers.
 
Additional Manufacturer Narrative
Correction made to d3 manufacturer email, g1 mfr site facility name, mfr site address 1, mfr site city, mfr site state, mfr site zip, mfr site country, mfr site phone, mfr first and last name, mfr email the product was not returned; therefore, no physical analysis of the product could be performed.Review of a photo of the device submitted showed that the fiber connector was detached from the fiber, confirming the reported clinical observations.Review of the device history record found no deviations that could have contributed to the reported event.Based on the available information, it is likely that a misaligned aiming beam caused the laser energy to heat an incorrect target ultimately leading to the detachment.Due to this, the most probable cause of cause traced to component failure has been applied to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LUMENIS SLIMLINE 200 D/F/L
Common Device Name
POWERED LASER SURGICAL INSTRUMENT
Manufacturer (Section D)
LUMENIS LTD
hakidma st 6
yokneam 20692 04
IS  2069204
Manufacturer (Section G)
ADVANCED FIBER TOOLS GMBH
bornheimer strasse 4
mittweida CA 9468
GM   9468
Manufacturer Contact
john stoffregen
4100 hamline avenue north
building c
saint paul, MN 55112
MDR Report Key15629101
Report Number2124215-2022-41953
Device Sequence Number12764333
Product Code GEX
UDI-Device Identifier07290109140261
UDI-Public07290109140261
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K170121
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/06/2024
Device Model Number1911-07
Device Catalogue Number1911-07
Device Lot Number0085800222
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/20/2022
Supplement Date Received by Manufacturer11/10/2022
09/21/2023
Initial Report FDA Received Date10/18/2022
Supplement Report FDA Received Date11/16/2022
10/01/2023
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-