| Model Number |
N/A |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/28/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer has received the sample and is pending evaluation.Results are expected soon.A lot history review (lhr) review is not possible, as no manufacturing lot number has been provided by the complainant.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer "on 9/28 our team reported that a safestep huber needle was found to be cracked and leaking while in use.There was no harm to the patient.".
|
| |
|
Additional Manufacturer Narrative
|
|
The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of a leaking infusion set was confirmed and the cause is currently under investigation.The product returned for evaluation was one 19ga x 0.75" safestep safety infusion set.The returned product sample was evaluated and the following observations were made: ¿ a tear in the extension tube was seen at the interface of the proximal luer adapter the fracture surfaces of the damage contained striation-like patterns, radiating tear marks and material buckling which were indicative of flexural fatigue based failures, which may have been a contributing factor to the extension tubing damage.Repetitive mechanical stresses such as twisting and kinking may have contributed to the observed event; however, it appeared that additional unidentified factors also contributed.The device is a supplied component and the supplier has been notified of this event.
|
| |
|
Event or Problem Description
|
|
It was reported by the customer "on 9/28 our team reported that a safestep huber needle was found to be cracked and leaking while in use.There was no harm to the patient.".
|
| |
|