| Model Number |
G408321 |
| Medical Device Problem Code |
Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Innovative health, llc became aware on (b)(6) 2022 of a report from marin general on a agilis nxt steerable introducer that was reported to not work properly.It is unknown when the issue occurred and when it was identified.Innovative health received the device for evaluation on (b)(6) 2022.Upon investigation, the side-port was identified to be detached.No injury was reported.
|
| |
|
Event or Problem Description
|
|
This device was reported to not work properly.Upon inspection, the device was found to have a detached side-port.It is unknown when the issue was identified.No patient injuries were reported.
|
| |
|