• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INNOVATIVE HEALTH, LLC.; STEERABLE INTRODUCER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INNOVATIVE HEALTH, LLC.; STEERABLE INTRODUCER Back to Search Results
Model Number G408321
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/01/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Innovative health, llc became aware on (b)(6) 2022 of a report from marin general on a agilis nxt steerable introducer that was reported to not work properly.It is unknown when the issue occurred and when it was identified.Innovative health received the device for evaluation on (b)(6) 2022.Upon investigation, the side-port was identified to be detached.No injury was reported.
 
Event or Problem Description
This device was reported to not work properly.Upon inspection, the device was found to have a detached side-port.It is unknown when the issue was identified.No patient injuries were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NA
Common Device Name
STEERABLE INTRODUCER
Manufacturer (Section D)
INNOVATIVE HEALTH, LLC.
1435 north hayden road
suite 100
scottsdale AZ 85257
Manufacturer (Section G)
INNOVATIVE HEALTH, LLC.
1435 north hayden road
suite 100
scottsdale AZ 85257
Manufacturer Contact
giancarlo montiel
1435 north hayden road
suite 100
scottsdale, AZ 85257
MDR Report Key15629434
Report Number3011610434-2022-00012
Device Sequence Number7941169
Product Code PNE
UDI-Device Identifier10841898128196
UDI-Public10841898128196
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K170311
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Nurse
Type of Report Initial
Report Date (Section B) 09/21/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/19/2023
Device Model NumberG408321
Device Catalogue NumberG408321
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2022
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/18/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured05/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-