A batch review of all the reported finished good lots and components confirmed there were no quality issues noted throughout the incoming inspection, manufacturing, sterilization, in-process, or final inspection process.No samples or photos of the surgery sites were returned for review.Retained samples were not available for review.If samples, photos or additional information becomes available at a later date, the samples/photos or information will be reviewed and added to the file and a follow-up report will be filed.Without receiving sterile devices from the reported lots to review/test, receiving photos of the surgical site or receiving detailed information regarding the pre-operative preparation of the devices, placement of the device in the tissue, method utilized to secure the device in the tissue as stated in the ifu, details regarding the surgeon¿s training, experience and/or technique with the device(s) or robotic surgery, the patient¿s health status and specifics, quality of the tissue, post-operative events that may have occurred that may have affected the outcome of the surgery, a definitive root cause for the reported event cannot be confirmed with certainty.As stated in the ifu for the devices, ¿adverse events including wound dehiscence and reactions are common risks/complications of any surgical procedure.There are many causes that can result in the wound opening, sutures failing or reaction, infection, abscess/leakage during or post-operative a procedure: the patient¿s health status, poor skin quality, thin skin; the risk is higher with a patient with a weak immune system, malnutrition or chronic medical condition.The surgical procedure ¿ the risk increases with poor surgical techniques such as improper suturing, over-tightened sutures or inappropriate type of suture used for a particular procedure.Other factors - the risk is greater with smoking, obesity, premature post-surgery exercise and heavy lifting.The warning in the ifu states, ¿this an absorbable material, the use of supplemental non absorbable sutures should be considered by the surgeon in the closure of sites which may undergo expansion, stretching or distention or which may require additional support.The ethicon sales representative submitted all the stocked lots in the operating rooms.Below are the possible lot numbers used, the manufacture dates, expiration dates and udi information as reported to surgical specialties by the ethicon sales representative: lot #: bb05ukl, manufacture date: 2021-11-05, expiration date: 2026-11-04, udi: (b)(4).Lot #: c413ylc, manufacture date: 2022-04-16, expiration date: 2022-4-15, udi: (b)(4).Lot #: c605hoa, manufacture date: 2022-06-05, expiration date: 2027-06-04, udi: (b)(4).
|
|
The actual device was not returned for review.No retained samples or photos are available for testing/review.If samples/photos become available, they will be reviewed, and the results will be included in the file.A revised response will be forwarded to you at that time.Ten (10) sterile samples from the finished good lot c605hoa were received and visually evaluated.No defects or abnormalities were observed on the packages or device(s).The samples met specification including the tensile requirements for a monoderm devices.Four (4) sterile samples from the finished good lot c413ylc were received and visually evaluated.No defects or abnormalities were observed on the packages or device(s).The samples met specification including the tensile requirements for a monoderm devices.Without receiving photos of the surgical site/device or receiving detailed information regarding the pre-operative preparation of the devices, placement of the device in the tissue, method utilized to secure the device in the tissue as stated in the ifu, the patient¿s specifics, quality of the tissue, post-operative events that may have occurred that may have affected the outcome of the surgery , details regarding the training, experience or technique of the surgeon with the use of the barbed device(s) or robotic surgeries, a definitive root cause for the reported event cannot be confirmed with certainty.
|