MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 4.5 ACCOLADE (127 DEG) SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH ACCOLADE (127 DEG) SIZE 4.5 ACCOLADE (127 DEG) SIZE 4.5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 6021-4535
Medical Device Problem Codes Degraded (1153); Detachment of Device or Device Component (2907); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 08/15/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Distributor reported that "patient required a total revision of right hip.".
 
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
 
Additional Manufacturer Narrative
An event regarding disassociation involving a lfit v40 cocr head that was mated with an accolade stem was reported.The event was confirmed via evaluation of the returned devices.Method & results -product evaluation and results: product evaluation and results: visual inspection: visual inspection of the returned devices indicated that the trunnion of the stem is severely worn.The metal head has minor damage from explantation.Material analysis: damage consistent with the loss of taper lock was observed on the head and stem of the returned devices.This has been documented as part of capa.No further material analyses were performed.-clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the reported accolade stem was mated with a lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Distributor reported that "patient required a total revision of right hip." update 11/15/2022: based on product inspection of the returned device, loss of taper lock was noted.
 
Event or Problem Description
Distributor reported that "patient required a total revision of right hip." update 11/15/2022: based on product inspection of the returned device, loss of taper lock was noted.(b)(6) 2023 - med review states: 'on (b)(6) 2022, the patient underwent a revision at which time black fluid and black staining of the hip was noted.The femoral head was loose on the trunnion and the trunnion was eroded.Elevated ion levels were present.The femoral stem was well fixed but there was bone loss present.It was decided to delay reconstruction and a spacer was implanted.'.
 
Additional Manufacturer Narrative
An event regarding disassociation, wear and abnormal ion levels involving a lfit v40 cocr head that was mated with an accolade stem was reported.The event was confirmed via evaluation of the returned devices and through a review of the provided medical records by a clinical consultant.Method & results -product evaluation and results: product evaluation and results: visual inspection: visual inspection of the returned devices indicated that the trunnion of the stem is severely worn.The metal head has minor damage from explantation.Material analysis: damage consistent with the loss of taper lock was observed on the head and stem of the returned devices.This has been documented as part of capa.No further material analyses were performed.-clinician review: a review of the provided medical records by a clinical consultant indicated: 'this patient underwent a right primary total hip replacement at some point in the past.He developed symptoms related to trunnion deformity with loosening of the femoral head with metallosis and bone loss with elevated metal ion levels.A two-stage revision was carried out initially to debride the hip, inserting a spacer and then at a second operation the hip was reconstructed.I can confirm that the first revision and the second revision occurred since i was able to review the operation reports.I cannot determine the root cause of this event with certainty.The causes of trunnionosis and metallosis elevated metal ion levels are multifactorial including surgical technique factors, patient activity level and bmi, and implant factors.The explanted prosthesis should be returned to stryker for examination and analysis by engineers.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock on with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the reported accolade stem was mated with a lfit v40 cocr head.The cocr head has been identified to be within scope of nc and capa.Visual inspection of the returned devices indicated that the trunnion of the stem is severely worn.The metal head has minor damage from explantation.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCOLADE (127 DEG) SIZE 4.5 ACCOLADE (127 DEG) SIZE 4.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
joann ripoli
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key15630876
Report Number0002249697-2022-01517
Device Sequence Number20323709
Product Code LPH
UDI-Device Identifier04546540510761
UDI-Public04546540510761
Combination Product (Y/N)N
Initial Reporter CountryEI
PMA/510(K) Number
K121308
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/31/2015
Device Model Number6021-4535
Device Catalogue Number6021-4535
Device Lot Number34391902
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2022
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer11/15/2022
07/03/2023
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date12/09/2022
07/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/11/2010
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age74 YR
Patient SexMale
-
-