• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; BTR TUBE, TRACHEAL (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE FROVA INTUBATING INTRODUCER; BTR TUBE, TRACHEAL (W/WO CONNECTOR) Back to Search Results
Model Number G48303
Medical Device Problem Code Component Incompatible (1108)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer ref# (b)(4).Initial reporter occupation: purchasing portfolio manager.Similar to device marketed under pma/510(k): k161813.Investigation is still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event or Problem Description
According to initial reporter: piece of plastic sheared off when tube placed over, ending up in patient's airway.
 
Event or Problem Description
No additional information regarding the patient and/or event has been received since the previous medwatch report was sent.
 
Additional Manufacturer Narrative
Manufacturers ref# (b)(4).Summary of investigational findings: a scrape off from the catheter ended up in patient¿s airways when a tube was placed over.The scrape off was retrieved and did not affect the patient outcome.The frova introducer and the scrape off from the frova were returned for device evaluation.The frova introducer was returned kinked.A scrape mark was found on the introducer near the proximal end.Squeeze marks and an indentation were observed near the proximal end.The tip was without any observed nonconformances.It was assessed that because any discovered non-conformances were properly dispositioned before final release, there is evidence that the dhr was fully executed.In addition, no other lot related complaints have been received from the field.It is therefore concluded that there is no evidence that nonconforming product exists in house or in field.There are adequate controls in place to ensure the device was manufactured to specifications.According to the instructions for use the 14.0 french frova introducer has been designed for placement of a single lumen endotracheal tube whose inner diameter is 6 mm or larger.Furthermore, the ifu warn that the tube and the introducer must be lubricated before use and that care must be taken when introducing/removing the introducer from the endotracheal tube, as contact with sharp edges on the internal surface of the endotracheal tube may cause small fragments to be shaved off the catheter introducer during introduction/removal.Despite several attempts no information as to the use of a single or a double lumen endotracheal tube could be obtained.Therefore, based on the provided information and the device evaluation an exact cause for this event cannot be established.Cook medical will continue to monitor for similar events.This report is required by the fda under 21cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FROVA INTUBATING INTRODUCER
Common Device Name
BTR TUBE, TRACHEAL (W/WO CONNECTOR)
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
MDR Report Key15630941
Report Number3002808486-2022-00994
Device Sequence Number20825707
Product Code BTR
UDI-Device Identifier00827002483033
UDI-Public(01)00827002483033(17)221010(10)E3908347
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/18/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/10/2022
Device Model NumberG48303
Device Catalogue NumberC-CAE-14.0-70-FIC
Device Lot NumberE3908347
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/04/2022
Supplement Date Received by Manufacturer12/21/2023
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date01/18/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/10/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-