• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; INTRAOCULAR LENS Back to Search Results
Model Number DAT315
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/19/2022
Type of Reportable Event Malfunction
Event or Problem Description
A facility representative reported that during an intraocular lens (iol) implant procedure, a linear crack was observed after the insertion of iol was in to the patients eye.The surgeon removed the iol a replaced during the initial procedure.Additional information has been received and stated that the lens had a crack in it, not evident until implantation.No patient harm was reported.
 
Additional Manufacturer Narrative
A sample device was not returned for analysis.Complaint history and product history records were reviewed and documentation indicated the product met release criteria.Root cause has not been identified.There have been one other complaints reported in the lot number.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
Only a portion of the optic was returned for evaluation in the lens case.Solution was dried on the returned optic portion.The optic was cut and has multiple small cracks protruding from the cut damaged area.The observed damage is typical of insertion and removal.The file indicated the use of a qualified cartridge and handpiece.The file indicated the use of three viscoelastics.Two are qualified for the lens/cartridge combination used.The other viscoelastic is not qualified for the lens/cartridge combination used.The root cause for the reported complaint appears to be related to failure to follow the ifu (instructions for use).The file indicated the use of a viscoelastic that is not qualified for the product combination used.The ifu instructs that an company qualified delivery system and viscoelastic combination should be used.The use of an unqualified combination may cause damage to the lens and potential complications during the implantation process.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACRYSOF IQ VIVITY TORIC EXTENDED VISION UV ABSORBING IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15631194
Report Number1119421-2022-02217
Device Sequence Number15439014
Product Code POE
UDI-Device Identifier00380652427142
UDI-Public00380652427142
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P930014
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 11/22/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/08/2025
Device Model NumberDAT315
Device Lot Number15065285
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/23/2022
Supplement Date Received by Manufacturer10/27/2022
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date11/22/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC VISCOELASTIC SYSTEM; HEALON 5 VISCOELASTIC; MONARCH III IOL DELIVERY SYSTEM, CARTRIDGE D; MONARCH III IOL DELIVERY SYSTEM, INJECTOR
Patient Age63 YR
Patient SexFemale
-
-