Model Number CCWTT3
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Product history records were reviewed and documentation indicated the product met release criteria.Qualified associated products were indicated.The product investigation could not identify a root cause for the reported complaint.The product was not returned for evaluation.No determination can be made without physical evaluation of the complaint sample.The instruction for use (ifu) instructs to completely fill the cartridge with ophthalmic viscosurgical devices (ovds) (diagram provided) immediately prior to loading and delivery of the lens.Do not attempt to load the lens without adequate ovd in the device.Not adequately filling the device with viscoelastic will result in inadequate coverage of lens and the lens fold path with ovd, which may result in damage.Follow the section regarding directions for use for information on the maximum allowed time for the iol to stay in the folded condition.Failure to adhere to manufacturer¿s recommendations may result in iol damage.Ifu note: during lens loading and insertion, do not allow the company toric trifocal iol to remain in a folded condition within the selected iol delivery system for more than 3 minutes prior to completing insertion into the capsular bag.No other complaints reported in the lot.The manufacturer internal reference number is: (b)(4).
 
Event or Problem Description
A non-healthcare professional reported that during the intraocular lens implant procedure smudges found to be left on the lenses.Additional information has been requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON PANOPTIX TORIC TRIFOCAL UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key15631288
Report Number1119421-2022-02223
Device Sequence Number5810446
Product Code MJP
UDI-Device Identifier00380652457439
UDI-Public00380652457439
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2022
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/15/2025
Device Model NumberCCWTT3
Device Catalogue NumberCCWTT3.225
Device Lot Number15288175
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/21/2022
Initial Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
BSS STERILE IRRIGATING SOLUTION; MONARCH CARTRIDGE D; MONARCH INJECTOR; PROVISC OPHTHALMIC VISCOSURGICAL DEVICE
Patient SexUnknown
-
-