| Model Number |
115-222 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Patient reported "change in size of right chest" and later reported right rupture.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Explant is in the future.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: rupture.
|
| |
|
Event or Problem Description
|
|
Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.Additional, changed, and/or corrected data: b5, b6, d.1, d.2, d.4, h.4, d.6a, d.6b, g4, h4, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: based on the device analysis grid, the assessments of the complaint are: rupture: observed an opening assessed as an unidentified (tear) opening (shell thickness whiting to spec).Additional observations: no other observations observed.
|
| |
|
Event or Problem Description
|
|
Patient reported "change in size of right chest" and later reported right rupture.Device has been explanted and replaced.
|
| |
|