• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number H7493941832300
Medical Device Problem Codes Difficult to Remove (1528); Failure to Advance (2524); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 09/30/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that stent dislodgment occurred.The patient underwent computed tomography (ct) turn down complex multivessel percutaneous coronary intervention.The target lesion was located in the left anterior descending artery.A 3.00 x 32mm synergy xd drug-eluting stent (des) was advanced but failed to cross the lesion.During withdrawal, the stent was dislodged off the balloon in the radial artery and the entire system was pulled from radial access.The patient felt discomfort and pain.Eventually, a 2.5 x 24 synergy xd/ 3.0 x 38 synergy xd and a 4.5 x 16 synergy xd des were used to treat multivessel disease and the procedure was completed with a 3.0 x 38 synergy xd des.The patient was doing well and fully recovered post-procedure.
 
Event or Problem Description
It was reported that stent dislodgment occurred.The patient underwent computed tomography (ct) turn down complex multivessel percutaneous coronary intervention.The target lesion was located in the left anterior descending artery.A 3.00 x 32mm synergy xd drug-eluting stent (des) was advanced but failed to cross the lesion.During withdrawal, the stent was dislodged off the balloon in the radial artery and the entire system was pulled from radial access.The patient felt discomfort and pain.Eventually, a 2.5 x 24 synergy xd/ 3.0 x 38 synergy xd and a 4.5 x 16 synergy xd des were used to treat multivessel disease and the procedure was completed with a 3.0 x 38 synergy xd des.The patient was doing well and fully recovered post-procedure.It was further reported that the target lesion was calcified and tortuous.Additionally, radial access issues were encountered before and after the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key15631616
Report Number2124215-2022-41392
Device Sequence Number12240212
Product Code NIQ
UDI-Device Identifier08714729981008
UDI-Public08714729981008
Combination Product (Y/N)Y
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/26/2023
Device Model NumberH7493941832300
Device Lot Number0028311656
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/05/2022
Supplement Date Received by Manufacturer10/19/2022
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date11/04/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured10/26/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age63 YR
Patient SexFemale
-
-