| Catalog Number |
2J8003 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Event date: this event occurred during september 09, 2022 to september 23, 2022.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that four (4) non-dehp fluid path intravia containers 500 ml capacity leaked at the middle port seam during filling.There was no patient involvement.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: the actual devices were not available; however, five (5) companion sample were received for evaluation.A visual inspection was done which observed a roughness in the injection port; however, the other components are placed according to the specification.The issue of leak in seam was not observed during the sample evaluation.Therefore, the reported condition was not verified.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|