• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-RM/LL-SZ 5; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MCK TIBIAL BASEPLATE-RM/LL-SZ 5; PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM Back to Search Results
Model Number 180615
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes Ambulation or Postural Difficulties (2544); Insufficient Information (4580)
Date of Event 09/26/2022
Type of Reportable Event Serious Injury
Event or Problem Description
Tibial baseplate on restoris mck medial uni has cracked in situ.Original operation date friday (b)(6) 2021.Issue noticed mid-september by surgeon on x-ray and reported to rep.Implant was a size 5 rm/ll onlay tibial baseplate.Plan is to pull out implant and revise to total knee.
 
Additional Manufacturer Narrative
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted when additional information becomes available.Device not returned.
 
Event or Problem Description
Tibial baseplate on restoris mck medial uni has cracked in situ.Original operation date (b)(6) 2021.Issue noticed mid-(b)(6) by surgeon on x-ray and reported to rep.Implant was a size 5 rm/ll onlay tibial baseplate.Plan is to pull out implant and revise to total knee.
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a mako baseplate was reported.The event was confirmed via clinician review.Method & results: product evaluation and results: visual inspection, material analysis, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: a review of medical records with a clinical consultant indicated "review of the lateral view x-ray of the unicondylar arthroplasty demonstrates a vertical fracture of the mid portion of the tibial tray.There is no evidence of component loosening or ostelysis.Material failure of the tibial portion of a unicondylar arthroplasty is confirmed.The root cause of this failure cannot be determined from the limited amount of information provided for review.".Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: a review of medical records with a clinical consultant indicated "review of the lateral view x-ray of the unicondylar arthroplasty demonstrates a vertical fracture of the mid portion of the tibial tray.There is no evidence of component loosening or ostelysis.Material failure of the tibial portion of a unicondylar arthroplasty is confirmed.The root cause of this failure cannot be determined from the limited amount of information provided for review." no further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.Product surveillance will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MCK TIBIAL BASEPLATE-RM/LL-SZ 5
Common Device Name
PROSTHESIS, KNEE PATELLOFEMOROTIBIAL, PARTIAL, SEMI-CONSTRAINED, CEMENTED, POLYM
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
perla zima
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key15632145
Report Number3005985723-2022-00133
Device Sequence Number17638064
Product Code NPJ
UDI-Device Identifier00848486000783
UDI-Public00848486000783
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K090763
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2022
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/29/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number180615
Device Catalogue Number180615
Device Lot Number26170919-01
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/26/2022
Supplement Date Received by Manufacturer12/14/2022
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date12/29/2022
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/13/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age60 YR
Patient SexMale
Patient Weight140 KG
-
-