Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: three samples (model #20159e) were returned by the customer.It was reported that the set was not flushing.The sets were examined for defects and abnormalities.No defects or abnormalities were observed.The samples were connected to a cut off syringe tip to check that the fluid path is in an activated position/open when connected to a syringe.The samples were then attempted to be flushed with water using a 10ml bd syringe to confirm an open fluid path.The fluid paths of all samples were open and those samples were able to be flushed.The failure was unable to be replicated, and the complaint could not be verified.A device history record review for model 20159e lot number 21119807 was performed.The search showed that a total of (b)(4) units in 1 lot number was built on (b)(4) 2021.There were no quality notifications issued for the failure mode reported by the customer during the production build of this set.A root cause was unable to be determined because the failure was unable to be replicated.Due to an increase in incidents of this failure mode, a trend for this occlusion issue has been identified for this product line, a capa (corrective action preventative action) has been initiated, and a team has been assembled in order to investigate the issue.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|