| Model Number |
M0031441890 |
| Medical Device Problem Codes |
Defective Component (2292); Component Missing (2306)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/20/2022
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device is not available to the manufacturer.
|
| |
|
Event or Problem Description
|
|
It was reported that when the subject microcatheter was used in a patient with abdominal aortic aneurysm, the markers could not be seen under fluoroscopy.The procedure was completed successfully by replacing the subject device.No clinical consequences were reported to the patient due to this event.
|
| |
|
Event or Problem Description
|
|
It was reported that when the subject microcatheter was used in a patient with abdominal aortic aneurysm, the markers could not be seen under fluoroscopy.The procedure was completed successfully by replacing the subject device.No clinical consequences were reported to the patient due to this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Due to the automated manufacturing execution system (mes) there are controls in the manufacturing process to ensure the product met specifications upon release.The subject device is not available; therefore, visual testing as well as functional testing cannot be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.The reported complaint could not be confirmed and it could not be definitively determined if the device failed to meet specifications because the product was not returned.Additional information provided by the customer indicated that the device was prepared for use as per the directions for use, there was no damage noted to the packaging prior to opening the packaging, the device was confirmed to be in good condition during preparation/prior to use on the patient and continuous flush was set up and maintained throughout the clinical procedure.While there are a number of potential causes for the reported issue, because review and analysis of available information failed to identify a definitive cause, a cause of undeterminable was assigned.
|
| |
|