• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ANGIODYNAMICS OPS FIBER 55CM SPOTLIGHT; VENACURE ENDOVENOUS LASER TREATMENT FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIODYNAMICS OPS FIBER 55CM SPOTLIGHT; VENACURE ENDOVENOUS LASER TREATMENT FIBER Back to Search Results
Catalog Number EVLT55OPS
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/04/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation, however, the customer did provide a picture showing distal portion of fiber inside patient with fiber fracture outside the patient.The fiber was still in one piece, fractured but not detached into two pieces.The customer's reported complaint description of fiber fractured outside of the patient was confirmed based on evaluation of picture provided by the end user.Although the complaint was confirmed via provided picture, without receiving a fiber sample for evaluation a definitive root cause cannot be determined.A potential root cause is handling damage to fiber during use of the device.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.At the vendor facility, during the manufacturing process, fibers are repeatedly bent and coiled and there is a final visual inspection performed with a hene laser to check the entire fiber length for defects prior to shipment.The directions for use which is supplied to the end user with this catalog number contains the following statement "prior to and during use, avoid damaging the fiber by striking, stressing or excessive bending.Do not coil the fiber tighter than a radius of 60mm".A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event or Problem Description
A physician reported an issue with a fiber from an evlt kit with spotlight ops sheath 55cm.During a procedure, a section of the fiber, that was located outside of the patient, snapped while emitting laser energy.The fracture was closer to the hub than the tip of the fiber.The fiber was removed and a new ops laser fiber was used to complete the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPS FIBER 55CM SPOTLIGHT
Common Device Name
VENACURE ENDOVENOUS LASER TREATMENT FIBER
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587805
MDR Report Key15633232
Report Number1319211-2022-00044
Device Sequence Number5911651
Product Code GEX
UDI-Device IdentifierH787EVLT55OPS5
UDI-PublicH787EVLT55OPS5
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K041957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/19/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberEVLT55OPS
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/04/2022
Initial Report FDA Received Date10/19/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexFemale
-
-