| Model Number |
3213182 |
| Medical Device Problem Code |
Device Misassembled During Manufacturing /Shipping (2912)
|
| Health Effect - Clinical Code |
Bruise/Contusion (1754)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
End user reports his catheters "have the fin not in line with the coude tip on catheter and have caused bleeding with use." he states "he first started to notice this defect last month after he had bleeding with use of this catheter and he then started to examine the catheter itself prior to use.He said it has noticed this defect ever since he started examining his catheter prior to use and said "out of 15 - 5 or 6 are not right." it is not the mu that is typically affected.He always applies lubricant to the catheters prior to use.He caths through his urethra to empty his bladder from retention as he said his bladder doesn't work well as well as a "cystosopy" revealed he has many "pouch" areas in his bladder.He caths 3 x a day.He said he noted some of the catheters that have the misaligned coude tip to the fin are off ranging anywhere from 0-90 degrees when looking at the fin at the tip mostly with that coude being off the right and he said he only found one that bent about 25 degrees to the left compared to the fin.He is aware of proper coude placement and states when he uses these catheters that do not have the defect they work wonderfully.He said a few months he initially had bleeding with use of one of these he didn't have his glasses on and he went to use it, and noted a small amount of blood after cathing.He thinks it could have had this defect.Bleeding stopped immediately without intervention.He is not on blood thinners.He called his md and she advised him to observe and let her know if bleeding reoccurred at next cath.He said he continued to use these catheters and noticed no problems.He said the last time he had bleeding was last month and then he was able to examine the catheter and noted this defect.He notes because of this defect he can have difficulty passing it through his prostate into his bladder so he moves it slightly around until he can get it through.He said this isn't painful but can feel tender.Currently he said he tries to work being conscious of the defect to avoid bleeding and to keep that coude tip upwards by observing it prior and with feel of it in urethra.".
|
| |
|
Additional Manufacturer Narrative
|
|
Device 2 of 6.Based on the available information, this event is deemed to be a reportable malfunction.To date no additional information has been received.Should additional information become available, a follow-up report will be submitted.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Batch record review was conducted resulting in following: urinary catheter in question was manufactured under sap material id1706974, gentlecath ic item ch16x405mm(1x100pk)u and manufacturing lot # 2c00354 in amount 86 400 pcs in february 2022.Urinary catheter in question was manufactured under sub assembly lots #1l03600 and 2a04395 on machine a097.Lot 1l03600 was produced on december 12-14, 2021; lot 2a04395 on february 06-08 2022.According to the process instructions g805311 the position of connector indicator against bent tip of catheter is checked by technician during order set up- 2pcs from each moulding station and visual inspection with focus on tiemman indicator position was carried out by operators at the beginning of each shift according to tm-422.Review of the dhr showed that all relevant tests required during the manufacturing process, packaging process and final product release had been fulfilled and met the requirements.No nonconformity has been registered during the manufacturing process of the mentioned lots.The batch record review indicates no discrepancies.Machine a097 is equipped by vision system.The key point for installation and setting vision system is to provide a control of all catheters focus on catheters failure : - eyelet no punch - hangnail left in eyelets - position of connector indicator and bent tip of catheter vision system has to reject every catheter with at least of one failures mentioned above.Validation of the position of tiemann indicator was in scope ccr-01239.Ccr-01239 was sign on december 19th 2019.Lot in question was produced after validation of the camera.A query was run against product number gentlecath ic tiem for malfunction code uca-pmc07.04 which yielded no other occurrence for past 24 months.The issue occurrence is considered as isolated.No further action is required.Should additional information become available, a follow-up report will be submitted.Fda registration number reporting site: 1049092 manufacturing site: 3005778470.
|
| |
|
Event or Problem Description
|
|
To date no additional patient or event details have been received.
|
| |
|
Search Alerts/Recalls
|