MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL HD LAPAROSCOPE,EP,10MM,30 DEG X 330MM; RIGID ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL HD LAPAROSCOPE,EP,10MM,30 DEG X 330MM; RIGID ENDOSCOPE Back to Search Results
Model Number 242636
Medical Device Problem Codes Crack (1135); Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/03/2022
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Additional narrative: udi: (b)(4).Reporter is a j&j sales representative.The device manufacture date is currently unavailable.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.
 
Event or Problem Description
It was reported by the sales rep that preoperatively to an unknown surgery on (b)(6)2022 , it was observed that the hd laparoscope,ep,10mm,30 deg x 330mm device had a small fracture on it that was visible on the display screen.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: investigation summary: the complaint device is not being returned, therefore unavailable for a physical evaluation, however the customer provided a photo of the product identification numbers which contains the product code, the number was verified, and it's correct.No more photos were provided.As no defect was found, a potential cause cannot be associated to manufacturing, therefore a manufacturing record evaluation is not required.  as part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection of the photo, and considering that the device was not physically evaluated, this complaint cannot be confirmed.The device is required for testing, the photo provided does not contain enough evidence to determine why the customer experienced the failure, hands on analysis should provide the required evidence to provide a root cause.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to monitor additional complaint information for potential safety signals through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD LAPAROSCOPE,EP,10MM,30 DEG X 330MM
Common Device Name
RIGID ENDOSCOPE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ  CH-2400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key15633748
Report Number1221934-2022-03172
Device Sequence Number14921019
Product Code GCJ
UDI-Device Identifier10886705030224
UDI-Public10886705030224
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K941541
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number242636
Device Catalogue Number242636
Device Lot Number1379345
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/03/2022
Supplement Date Received by Manufacturer10/26/2022
Initial Report FDA Received Date10/19/2022
Supplement Report FDA Received Date10/27/2022
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-